Multiple Input Analyzer

UDA2182 Universal Dual Analyzer

One of the most versatile analytical instruments to date.

The UDA2182 monitors and controls analytical process variables such as acidity or alkalinity (pH), Oxidation Reduction Potential (ORP), contacting conductivity and Dissolved Oxygen (DO) concentration. Applications of the UDA2182 include power water quality control, wastewater influent and effluent, pure water preparation, food and dairy and neutralisation.
✓ Mix-And-Match Input Card Design
✓ Menu-Driven Configuration
✓ High Resolution Graphic Backlit Display
✓ Ethernet Communications
✓ Wireless IR Interface
✓ Multiple Language Prompts
✓ Easy to Install, Operate and Maintain
✓ Accurate Measurement

Consen Engineering Pte. Ltd.

Tech Specifications

CasePlastic Enclosure CSA Type 4X (NEMA 4X)
DisplayLCD Dot Matrix, 128 x 64 dpi
Display Accuracy0.05% of Reading
Control capabilities/advance featuresPID Control, Ethernet/Modbus Communications, Pocket PC and Infrared Configuration, , High Purity Water Solution Compensation, 0.2 sec Update Rate, E. European Languages
Operating Conditions0° to 60°C (32° to 140°F)
Operating Voltage90-264 Vdc; 47-63 Hz
Analogue OutputsUp to Three 4 to 20mA
RelaysUp to 4 Relays
MountingsPipe, Wall, or Panel
ApprovalsCE; FM Class 1, Div. 2; UL/CSA General Purpose

 

Download Tech Specs

pH/ORP Sensor

pH is measured in a wide variety of process applications to ensure product quality, monitor effluent discharge and to control addition of chemicals for optimum chemical reaction.

pH Input

The pH input in UDA2182 will accept a wide variety of sensors-non-glass Durafet®, HB high performance pH series and traditional glass Meredian® electrodes, ORP combination electrodes and the HPW700 high purity system. The pH input also features:
● Auto Buffer Calibration
● High Purity Water Solution Compensation
● 0.2 sec update rate for fast responding Durafet® pH Electrodes

Consen Engineering Pte. Ltd.

Tech Specification

7773 Mounting7777 Mounting
Measurement Range0-14pH
±1600 mV ORP
0-14pH
±1600 mV ORP
Temperature RangeDepends on SensorDepends on Sensor
Pressure and Temperature Ratings

Immersion/Polypropylene
689 kPa @ 60˚C
(100 psig @ 140˚F)

316 SS
689 kPa @ 80˚C
(100psig @ 176˚F)

Flow-through/polypropylene
689 kPa @ 60˚C
(150 psig @ 176˚F)

For Durafet II pH Electrodes 51205554-001(In-line) and Meredian II pH Electrodes:

345 kPa (50 psig) @ 100°C, 690 kPa (100 psig) @ 50°C

For ORP Meredian II electrodes 31074388 gold:

345 kPa (50 psig) @ 100°C, (690 kPa 100 psig) @ 50°C 31074389 platinum:

345 kPa (50 psig) @ 110°C, 690 kPa (100 psig) @ 50°C

Materials of Construction

Polypropylene, Ryton or 316 SS

Durafet and glass electrode bodies, Ryton

Mountings

Immersion or flow-through

Immersion or in-line tee (3/4 in. NPT fitting)

 

Download Tech Specs

 

Download Tech Specs

 

Conductivity Sensor

Measure conductivity, resistivity, salinity and chemical concentrations with Honeywell’s conductivity cells and mountings.

Conductivity Input

The conductivity input in UDA2182 indicators accepts signals from Honeywell’s standard selection of contacting conductivity cells. It features:
● Temperature Compensation Curves
● Calculation of % Rejection/Passage and difference of two cells
● Conversions to ppm, ppb or ppt Total Dissolved Solids (TDS)
● CO2, Concentration Algorithm
● pH from Differential Conductivity

Consen Engineering Pte. Ltd.
Sensors4973 Contacting Conductivity Cells4905 Contacting Conductivity Cells4909 Contacting Conductivity Cells5000TC Toroidal Conductivity Cells
Measurement Range0.01, 0.1, 1.0, 10 cell constants, 0.055μS/cm to 250 mS/cm0.01, 0.1, 1.0, 50 cell constants, 0.055μS/cm to 1S/cm0.01, 0.1, 10.0, 50 cell constants, 0.055μS/cm to 1S/cm0.2 to 200 milliSiemens/cm
Pressure and Temperature1724 kPa @ 140˚C (250 psig @ 284˚F)1034 kPa @ 130˚C (150 psig @ 266˚F)SS: 3.45 bar @ 140˚C (250psig @ 284˚f), CPVC 2.07 bar @ 140˚C (30psig @ 284˚FPolypropylene: 6.9 bar @ 100˚CC (100psig @ 212˚F), PVDF: 6.9bar @ 120˚C (100psig @ 248˚F), PEEK: 13.8 bar @ 150˚C (200psi @ 302˚F), PFA Teflon: 13.8 bar @ 150˚C (200psig @ 302˚F)
Material of ConstructionTitanium or graphiteNickel or platinumNickel or platinumPolypropylene, PVDF, PEEK, PFA Teflon
Mountings¾ inch NPT threaded fitting1-inch NPT threaded fittingInsertion/removal ball valve assembly in CPVC or SS allows insertion or removal of cell without stopping the processImmersion union adapter, sanitary 2-inch flange or insertion/removal

 

Learn More

 

Download Tech Specs

 

Dissolve Oxygen

Determine the levels of dissolved oxygen in water with a reliable and universal analytical indicator.

Companies use the universal UDA2182 indicator for its versatility in measuring conductivity, pH, ORP and dissolved oxygen. With its ‘mix-and-match’ input design, it offers unrivalled flexibility for a wide range of applications. Its common form, fit and function to older Honeywell analysers also enables a quick and easy retrofit into existing panels and installations.

Dissolved Oxygen Input

The dissolved oxygen input is from Honeywell’s unique equilibrium probe. It has these additional features:
● ppm or ppb Measurement
● Automatic or Manual Calibration
● Ambient Temperature and Atmospheric Pressure Compensation

Consen Engineering Pte. Ltd.

Greater value and enhanced performance

These analysers or probe systems determine the levels of dissolved oxygen in water. The patented equilibrium dissolved oxygen probe design is unaffected by inert fouling or changes in flow conditions. The system’s analyser/controller measures either ppb DO levels in power plant and semiconductor applications for corrosion detection or deaerator efficiency or ppm DO levels in wastewater, environmental and process applications for control and compliance.

DL5000 Equilibrium Probe for PPB & PPM Application

Measurement Range0-20,000 ppb or 0-20 ppm
Temperature Range2˚ to 60˚C (35.6˚ to 140˚F)
Pressure and Temperature Ratings

316SS: 50 psi (345 kPa)

CPVC: 30 psi (207 kPa)

Materials of Construction316SS or CPVC housing
Special FeaturesEquilibrium probe design requires no internal probe maintenance
MountingsImmersion in tank, in-line or sample flow chamber
Dimension (OD)219 x 34 mm (8.62 x 1.32 in), 1-inch NPT pipe size, 20-feet waterproof cable
Response Time85% in 60 seconds

 

Download Tech Specs

Gas Analyzer

Accurate Analysis of Binary Gases

The Honeywell 7866 Thermal Conductivity Analyzer is designed to provide a highly sensitive and accurate analysis of a binary (2- component) mixture of gases. It can be calibrated to measure a single component of a multicomponent gas mixture, providing the background gasses constitute a stable mixture (such as air), or have approximately the same thermal conductivity.

Thermal Conductivity

The 7866 Analyzer features a thermal conductivity system that measures concentrations of hydrogen purity and CO2 gas. This measurement is typically made in hydrogen-cooled generators.

  • Easy to use prompts
  • Security code-protected
  • Reliable solid state design
  • High response speed
  • High sensitivity
  • Excellent stability
  • Low maintenance requirement
  • Low installation costs through optional remote mounting capability of the sensing unit (transmitter)
  • Explosion-proof housing on the sensing unit available for Class1, Div1 areas
  • Signal transmission from the sensing unit up to 1000 feet over unshielded lead wires
  • Panel-mounted 1/4 DIN control unit with easy-to-read display
  • Current output signal from the control unit representing measured PV
  • Single or dual alarms
  • A triple range analyzer for hydrogen cooled generator applications
  • Optional Modbus communications supports configuration and data acquisition
Consen Engineering Pte. Ltd.
H2 Purity Gas Analyzer7866 Analyzer
Accuracy±2% of span
Response Time (for H2)Initial, <1 sec 63%, 13 sec, 90%, 23, 99%, 40 sec
Measuring Range1, 2 or 3 as specified
Sample Requirement (Sensing Unit)
0.2 to 4.2 cfh flow 37mm Hg Pressure min.
Power Requirement (Control Unit)
Universal 90 to 264 Vac, 50 to 60 Hz
Weight (Sensing Unit/Control Unit)
8.5 kg (18 3/4lb)/1.3kg (3.0 lbs)

 

Download Tech Specs

Interested? Get in touch.

Contact Us
Top